CSC AQUAPAL III COULAMETRIC KARL FISCHER TITRATOR
FEATURES
The  Aquapal III provides fast, easy, accurate results in PPM or percent on water in turbine, hydraulic, transformer, crude and many other oils. Its rugged design and simple one-button operation means the Aquapal can be operated by anyone in the plant to generate the quickest, fastest, most reliable results in a matter of minutes. 
Specifications

 Standard Specs:
 Range: 1 ppm to 100%
 Speed: 2 mg/minute
 Sensitivity: 0.1 ug
 Precision : 0.5% at 1mg
 Drift Correction:  Automatic
 Input/Output : Built-in RS232C for balance interface or computer
  
Unique Specs:
Single Reagent system available
Carry case offers portability for use in the field.
Compact size of 14”W x 8”D x 10.8”H or 350 x 200 x 280mm
Light weight of 15 lbs. or 6.8kg
Built in printer with standard paper backs up test with hard copy results.
Obtain results in percent solubility.
Simple menu operation
Run tests by weight, volume or dilution

Result Specs:
Result Format: ppm - % - ug
Data Calculation: mean, standard deviation and coeffiicient of variation
Memory: stores from 1 -99999 sample I.D. numbers
  
Other Specs:
Tests according to ASTM D4928, D1533, D1744 and equivalent methods.
CE approved manufacturer.
OPERATORS MANUAL
700 26100000 CSC Aquapal III Karl Fischer Moisture Analyzer
with carrying case
                                         OVER TITRATION

Some times the Aquapal can over titrate; this is not normally an indication of a fault, only if it continuously over-titrates. The purpose of this test is to establish if leaving the instrument on for a day and only injecting Hydranal water standards, does the instrument become unstable and keep going into excess drift, so lets go through the test. 

1.      With a clean cell and new reagent run the APIII for a day occasionally injecting Hydranal water standards, monitor and record the drift.

DO NOT INJECT your sample,  the reason for doing this is to see if a component in the electronics is breaking down after the instrument has been turned on for a period of time.

2.      If the Aquapal goes into over-titrating just add a Hydranal water standard to bring it out of over-titration. (If the instrument is stable after a day running water standards, this would indicate the instrument and glassware is okay and the problem lies elsewhere.) 

3.      When did the Problem start:
 During a run of test or after charging with new reagent. 
                          Reagent
■      recently switched to a different reagent?
■      Old reagent?

HYDRANAL Water Standard Test

●      Draw up more than 2ml
●      Push out air
●      Weigh syringe 
●      Type weight in
●      Inject 1ml into glassware (hit start while injecting)
●      Re-weigh 
●      Run for awhile.. it asks you to enter new weight.. 
●      Enter new weight
●      Print out results
        .1 + 10ppm 
        1.0 + 30ppm

Repeat three more times


SPECIAL REAGENTS AVAILABLE FOR 
KETONES AND ALDEHYDES

INITIAL PURCHASE INCLUDES STARTING SUPPLIES & REAGENTS
The AquaPal III produces iodine at the wire mesh of the generator electrode. When electricity is conducted across the mesh, one mole of the created iodine, consumes one mole of water. One milligram of water is equivalant to 10.71 coloumbs of electricity. The AquaPal measures the electrical current needed to create iodine, and remove the existing water.

Employing a stepped pulse current, the AquaPal III automatically selects titration speed depending on the amount of water present. As the endpoint is reached, the titration rate slows until the detector signal reaches and maintains the original baseline value. This endpoint is a minute excess of free iodine. At that time, the AquaPal III subtracts any background drift, then calculates and prints the water content.
SINGLE REAGENT OR DUAL REAGENT OPTION
Instructions Included for:
                                   SEND MESSAGE HERE
INSTANT QUOTATIONS INCLUDING SHIPPING, AND ONLINE ORDERING AVAILABLE AT 



FOR A QUOTATION , SELECT PURCHASE ORDER AS PAYMENT OPTION, AND TYPE "QUOTE" IN FIELD PROVIDED 
NO ITEMS WILL BE SHIPPED WITHOUT A VALID PO BY AN APPROVED COMPANY

GENERAL INFORMATION REQUEST
WEBSTORE
PRINTER ACCESSORY
EMAIL ADDRESS
NAME
LOCATION ZIPCODE
CALL BACK DATE
COMPANY
TELEPHONE
TIME (USE LOCAL TIME)
BROCHURE
CURRENT PRICING
Add this page to your favorites.
PACIFIC GAS AND ELECTRICBONNEVILLE POWER AUTHORITY
SOUTH CAROLINA ELECTRIC AND GAS    TENNESSEE VALLEY AUTHORITY
ALABAMA POWER COMPANY         BALTIMORE GAS AND ELECTRIC
IDAHO POWER COMPANY     DUKE POWER COMPANY
BUCKEYE PIPELINE     IMPERIAL OILS
LOUISVILLE GAS AND ELECTRIC    OHIO POWER COMPANY
PEPCO      TEXAS UTILITIES
SOUTH DAKOTA SOYBEAN PROCESSINGUTAH POWER
PECO ENERGY COMPANY    ABB POWER
NEW YORK STATE ELECTRIC & GAS        OHIO POWER COMPANY
SGS CONTROL SERVICES    BALTIMORE GAS & ELECTRIC
WISCONSIN ELC POWER COMPANY        WESTINGHOUSE ELECTRIC CORP
 EQUIPMENT FOR QUALITY CONTROL AND R&D APPLICATIONS
www.qclabequipment.com
Telephone    (804) 318-3686
Toll Free US (866) 244 1578
Fax               (703) 991-7133
Hours      8 AM - 6 PM EST
INTERNATIONAL & DOMESTIC SALES
CUSTOMER SERVICE * SALES * TECHNICAL SUPPORT

This page was last updated: March 3, 2012
MADE IN THE U.S.


WHITESTONE VA
TEL (804) 318-3686 FAX (703) 991-7133
INFORMATION@QCLABEQUIPMENT.COM 
WWW.QCLABEQUIPMENT.COM
ALL RIGHTS RESERVED QAQC